Abbott Amplatzer Talisman transcatheter PFO occluder
Transcatheter PFO closure device · Структурные вмешательства

Amplatzer™ Talisman™ PFO OccluderКод изделия (REF): 9-PFO-1818

Over 25 years ago, Abbott pioneered PFO closure with the Amplatzer™ PFO Occluder. With the introduction of the Amplatzer Talisman PFO Occluder, that proven Amplatzer PFO closure technology has been built upon: the occluder now comes assembled and ready to use, simplifying prep and enhancing ease of use. The 30 mm size has been refined to have a smaller left atrial disc to better avoid interference with surrounding structures. The occluder is offered in four sizes, from a right atrial / left atrial disc combination of 18 mm / 18 mm up to 35 mm / 25 mm, with recommended sheath sizes of 8 F or 9 F (45° curve). Device selection follows the published sizing guidelines: a 25 mm device is suggested for a simple PFO or PFO with a non-prominent ASA; a 30 or 35 mm device for complex PFO with anatomical characteristics such as an atrial septal aneurysm with a long tunnel or thickened septum secundum; and an 18 mm device for PFO with small anatomy (septal primum length < 20 mm). Abbott advises evaluating device position after deployment but before detachment, using echocardiography to ensure the device does not impinge on the free atrial wall or aortic root, and recapturing and redeploying if position is unsatisfactory.

Поставляется под каталожным кодами 9-PFO-1818, 9-PFO-2518, 9-PFO-3025, 9-PFO-3525, 9-TDS-08F45-80, 9-TDS-09F45-80, 9-GW-002.

Размеры и характеристики

Indication
Patent foramen ovale (PFO) closure
Disc combinations
18/18, 25/18, 30/25, 35/25 mm (RA/LA)
Recommended sheath
8 F (18/25 mm) or 9 F (30/35 mm), 45° curve
Delivery
Comes assembled and ready to use

Производитель: Abbott Laboratories · Источник данных: cardiovascular.abbott

Конструкция и технологии

Технические характеристики и схемы согласно публикациям Abbott.

Device profile

The refined 30 mm size features a smaller left atrial disc to better avoid interference with surrounding structures.

Official Abbott profile image of the Amplatzer Talisman PFO Occluder

Идентификация и характеристики изделия

Референсные идентификаторы для закупок, регистрации и тендерной документации.

Код изделия9-PFO-1818
Номера REF9-PFO-1818, 9-PFO-2518, 9-PFO-3025, 9-PFO-3525, 9-TDS-08F45-80, 9-TDS-09F45-80, 9-GW-002
ПроизводительAbbott Laboratories
БрендAbbott
Семейство изделийСтруктурные вмешательства
Доступность по странамТаджикистан, Афганистан

Клиническое применение

Amplatzer™ Talisman™ PFO Occluder — это transcatheter pfo closure device семейства «Структурные вмешательства» производства Abbott Laboratories, применяемое специалистами по интервенционным вмешательствам в процедурах, где показано transcatheter pfo closure device данного класса. Поставляется в 7 конфигурациях (каталожные коды с 9-PFO-1818 по 9-GW-002) для соответствия анатомическим и процедурным требованиям.

Полные показания к применению, противопоказания и инструкцию по применению (IFU) — официальную документацию Abbott — можно запросить у нашей клинической команды. Modern Medical Solutions поставляет Amplatzer™ Talisman™ PFO Occluder в больницы и катетеризационные лаборатории по всему Таджикистану и Афганистану.

Информация для заказа

Каталожные референсные коды согласно публикациям Abbott. Свяжитесь с нами, чтобы уточнить текущее наличие.

Amplatzer™ Talisman™ PFO Occluder

Model/Reorder Number9-PFO-1818
Right Atrial Disc Diameter (A)18 mm
Left Atrial Disc Diameter (B)18 mm
Recommended Sheath Size8 F; 45° curve
Model/Reorder Number9-PFO-2518
Right Atrial Disc Diameter (A)25 mm
Left Atrial Disc Diameter (B)18 mm
Recommended Sheath Size8 F; 45° curve
Model/Reorder Number9-PFO-3025
Right Atrial Disc Diameter (A)30 mm
Left Atrial Disc Diameter (B)25 mm
Recommended Sheath Size9 F; 45° curve
Model/Reorder Number9-PFO-3525
Right Atrial Disc Diameter (A)35 mm
Left Atrial Disc Diameter (B)25 mm
Recommended Sheath Size9 F; 45° curve

Delivery accessories

Model/Reorder Number9-TDS-08F45-80
ItemAmplatzer™ Talisman™ Delivery Sheath
Specifications8 F; 45° tip angle; sheath inner diameter 2.69 mm; sheath outer diameter 3.45 mm; usable length 80 cm
Model/Reorder Number9-TDS-09F45-80
ItemAmplatzer™ Talisman™ Delivery Sheath
Specifications9 F; 45° tip angle; sheath inner diameter 3.00 mm; sheath outer diameter 3.81 mm; usable length 80 cm
Model/Reorder Number9-GW-002
ItemAmplatzer™ Guidewire
Specifications0.035 inch diameter; Super Stiff body; 1.55 mm, Modified J-tip; usable length 260 cm

Model/Reorder numbers and dimensions per the official Abbott Amplatzer™ Talisman™ PFO Occluder product page ordering information. A = right atrial disc diameter, B = left atrial disc diameter.

Документация и материалы

Инструкцию по применению (IFU), технические спецификации и клиническую литературу по Amplatzer™ Talisman™ PFO Occluder можно запросить у нашей клинической команды.

Запросить документацию

Часто задаваемые вопросы

What is new about the Amplatzer™ Talisman™ PFO Occluder?
The Amplatzer Talisman PFO Occluder builds upon Abbott's proven Amplatzer PFO closure technology. It now comes assembled and ready to use, simplifying prep and enhancing ease of use, and the 30 mm size has been refined with a smaller left atrial disc to better avoid interference with surrounding structures.
What sizes and sheath sizes are available?
The occluder is offered in four sizes: 9-PFO-1818 (18/18 mm), 9-PFO-2518 (25/18 mm), 9-PFO-3025 (30/25 mm) and 9-PFO-3525 (35/25 mm), where the dimensions are right atrial disc / left atrial disc diameter. The 18 and 25 mm devices use an 8 F sheath and the 30 and 35 mm devices use a 9 F sheath, each with a 45° curve.
How is device size selected?
Per the published sizing guidelines, a 25 mm device is suggested for a simple PFO or PFO with a non-prominent ASA; a 30 or 35 mm device for complex PFO with characteristics such as an atrial septal aneurysm with a long tunnel or thickened septum secundum; and an 18 mm device for PFO with small anatomy (septal primum length < 20 mm).
What should be checked before device detachment?
Abbott advises evaluating the position of the device after deployment but before detachment, using echocardiography to ensure the device does not impinge on the free atrial wall or aortic root. If the device interferes with an adjacent structure, it should be recaptured and redeployed, or replaced with a smaller device.